|     Specification:  |        Value, Value Range,   or Quality:  |        Rationale:  |   
|     Counting   Time Cycle  |        1   Minute (30 seconds “On,” 30 Seconds “Off”)  |        After   turning on the machine, the patient has 30 seconds to prepare themselves. The   next 30 seconds is used for monitoring to allow for an accurate pulse while   not requiring the patient to stay for a full minute.  |   
|     Red   LED  |        Pulse   detection (Either On or Off)  |        The   LED is there to help the patient ensure that a correct reading is occurring.   If the light is blinking spastically, this signals to the patient that there   is either some interference or the patient is not remaining still. If the   light is not blinking at all then there is something wrong with the system as   this symbolizes no interruption in the detection of IR light.  |   
|     Serial   Monitor Output  |        Raw   values and “alpha value” corrected values  |        The   numbers displayed on the Serial monitor are there to allow for the user to   see the actual value changes from the IR sensor. This helps with   troubleshooting and also allows for the alpha value correction to smooth out   the number series in case graphing is needed.  |   
|     Serial   Baud Rate  |        57600   or 115200  |        In   order to successfully detect changes in IR light transmittance, the baud rate   for the serial monitor must be at one of the highest possible levels.  |   
|     Resistor   attached to the IR LED  |        2 kΩ  |        Using   a potentiometer, this resistance value was calculated to be the best resistor   value to allow for successful pulse detection.  |   
Welcome to our BE 470: Medical Devices blog. Over the Spring 2011 semester we will be bringing you updates about our progress in that class. This will include the circuitry we used, the programming, and probably even some pictures. Enjoy our blog!
Thursday, May 5, 2011
The Final Project: Specifications of our Device
Below is a list of specifications relating to our device. Each was chosen to provide the user with a better experience when using the product:
  